Analytical chemistry · Human-reviewed original
Question
A linear calibration has been checked only over 0.02–0.20 M. An unknown initially gives a signal corresponding to about 0.8 M by extrapolation. Which measurement plan best supports a reliable result within the validated range?
- Dilute a measured aliquot by a known factor and remeasure within the calibrated range.
- Discard all calibration points below the sample signal
- Report the extrapolated value as equally validated without further work
- Use an unknown dilution factor because only signal reduction matters
Show answer and explanation
Answer: Dilute a measured aliquot by a known factor and remeasure within the calibrated range.
The sample lies outside the interval over which the calibration has been checked. Choose a known dilution sufficient to place its signal inside that interval. Measure the diluted sample and use the dilution factor to recover the original concentration, preserving traceability rather than assuming distant extrapolation remains linear. Option review: A: This brings the signal into the validated interval and permits back-calculation by the known dilution. B: Removing standards does not validate the distant extrapolation. C: Linearity outside the checked interval has not been demonstrated. D: An unmeasured dilution prevents quantitative reconstruction of the original concentration.