ANALYTICAL CHEMISTRY · 8 MIN LESSON

Make the measurement model explicit

A concise core lesson, followed by independent practice.

Practice this topic (sign in required)
RELATED LECTURE · MIT OPENCOURSEWARE

Acid-Base Titrations Part I

A supporting university lecture. The written lesson and check can be used independently.

Loads YouTube and shares request information with Google.

MIT source & transcript · Catherine Drennan · 5.111, Fall 2014 · CC BY-NC-SA; source terms

THE CONCEPT

Translate the experimental signal into concentration or amount through a chemical or calibration model. Separate systematic bias from random uncertainty and preserve the precision supported by the measurement.

The core relationship

A=εlc,usum=u12+u22A=\varepsilon lc, \quad u_{sum}=\sqrt{u_1^2+u_2^2}
WORKED EXAMPLE

For A = 0.200c + 0.010 and measured A = 0.410, c = 2.00 in the calibration’s concentration units. Independent reading uncertainties of 0.03 and 0.04 mL combine to 0.05 mL.

The common trap

Subtract a calibration intercept before applying the slope. Repetition can reduce random noise but cannot by itself remove systematic bias.

Go deeper

Evaluate calibration range, blank correction, matrix effects and uncertainty propagation before trusting a fitted concentration.

The additional instruction below was developed with AI assistance and awaits independent educator review.

How to approach a problem

A calibration curve connects a measured signal to concentration over a stated working range. Correct for a blank only as specified by the method, then use the calibration relation and account for dilution. Measurements outside the calibrated range may require a new preparation or an appropriate validated method rather than blind extrapolation.

Another worked example

Suppose a blank-corrected calibration is A = 0.0200c, with c in mg L⁻¹, and the blank-corrected sample absorbance is 0.300. The measured solution has c = 15.0 mg L⁻¹. If it was prepared by a tenfold dilution, the original concentration is 150 mg L⁻¹, assuming the result is within the calibration range.

Check your reasoning

Do not subtract the blank twice. The calibration's slope units, intercept, range, and dilution factor all affect the result.

Try a reviewed sample

No account required. Sample results stay on this device.

SEL-0382

Retention time identifies elution order

A chromatogram has analyte peaks with retention times 2.0, 5.0 and 8.0 minutes. The injection is at time zero. What does the 2.0-minute peak represent?

Human-reviewed originalSolve this sample →
SEL-0383

A blank corrects a background signal

A reagent blank processed identically to samples gives absorbance 0.030. A sample gives 0.210, and the method uses additive blank correction. What analyte-associated absorbance is used?

Human-reviewed originalSolve this sample →
SEL-0386

Beer–Lambert scaling changes with path length

An ideal solution has absorbance 0.240 in a 1.00 cm cell at a fixed wavelength. It is transferred unchanged to a 2.50 cm cell. What absorbance follows within the linear Beer–Lambert regime?

Human-reviewed originalSolve this sample →
SEL-0387

Chromatographic areas need response-factor correction

Two analytes give integrated peak areas 120 for A and 80 for B. Their detector response factors are 2.0 area units per nmol for A and 1.0 for B. What mole fraction of A is represented?

Human-reviewed originalSolve this sample →
SEL-0388

Aliquot dilution must be carried back to the original sample

A 5.00 mL aliquot is diluted to 50.00 mL. The diluted sample is measured as 0.00200 M analyte. What original concentration follows if no analyte is lost?

Human-reviewed originalSolve this sample →
SEL-0389

A standard should bracket the sample signal

A linear calibration has been checked only over 0.02–0.20 M. An unknown initially gives a signal corresponding to about 0.8 M by extrapolation. Which measurement plan best supports a reliable result within the validated range?

Human-reviewed originalSolve this sample →
SEL-0391

Two wavelengths can resolve a two-component spectrum

An ideal 1 cm cell contains A and B. At two wavelengths, absorbances satisfy y₁=2c_A+c_B and y₂=c_A+3c_B, with concentrations in mM and the given coefficients in mM⁻¹. Measurements are y₁=0.50 and y₂=0.50. What concentrations follow?

Human-reviewed originalSolve this sample →